Ultrasonic Flow Meters

20+ Years Manufacturing Experience

What to do in case of sensor failure given that the sensors are paired with the transmitter according to the calibration procedure ?

If one of paired sensor is failed and cannot be repaired,
1. to changed another new paired ( 2pcs) sensors.
2. to send the work normal sensor to our factory to pair another one.
If the two sensors are not paired sensors, the meter can not work well and it will affect the meter accuracy.
 
If the transmitter is failed and can not be repaired,
It is ok to work with other paired sensor, it will have a slight effect on accuracy, negligible.
Of course, the best way is re-calibrate the transmitter and paired sensors, because it is a complete calibration process.

Post time: Nov-13-2023